Cm2600d spectrophotometer is a portable color measuring instrument produced by Japan. Users may encounter problems such as "the instrument cannot be connected with the data management software" and "system error" in the process of a long time. Here are solutions for you.
Common faults of cm2600d spectrophotometer:
1. What should be done if Japan color measuring instrument cannot be connected with data management software?
Under the premise of ensuring the integrity of hardware conditions such as connecting line and computer port, it is necessary to confirm whether the port number setting in the software matches the computer display (my computer ---- properties ---- Hardware ---- Device Manager --- port). In addition, if portable instrument is used (such as cm-2600d, cm-512m3, etc.), confirm that the instrument has been set to remote mode.
2. Japan cm-2600d switched the lens position switch and the target hood from MAV to sav, but the error of "ER25 please confirm the measurement system" appeared as soon as it was corrected.
In the "target / mask" of the environmental conditions ("cond1" ~ "cond6") to be used on cm-2600d, change the part of the target mask to s. Please refer to cm-2600d operation manual for operation method.
Precautions for use of cm2600d spectrophotometer:
Operating environment:
It should be used in the environment with external temperature ranging from 5 ℃ to 40 ℃, not in the environment with rapid temperature change.
Do not place the instrument directly in the sun or near the furnace.
Do not use the instrument in the environment of smoking, dust and chemical gas.
Do not use the instrument in a strong magnetic environment.
It is not allowed to use the instrument at an altitude of more than 2000m.
Cm-2600d / 2500d is an indoor test instrument and should not be used outdoors.
White correction plate:
Because the white board data is measured at 23 ℃, in order to get more accurate measurement data, white board correction and measurement should be carried out at about 23 ℃. Protect the whiteboard from scratches or contamination. Please cover the whiteboard when it is not in use.
Aperture: do not touch the inner surface of the aperture by hand, and do not scratch or stain it. When not in use, put it in the hole diameter storage place in the whiteboard correction box.
Power supply: be sure to turn off the power switch when the instrument is not in use. Always use a transformer (ac-a17) for power supply.
System: do not allow the instrument to be impacted or vibrated with great intensity. In order to make the integrating sphere and measuring hole of the instrument free from pollution, remember to place the instrument on the whiteboard calibration box when the instrument is not in use. It can't be used beside the TV, recorder and other electrical equipment, or the instrument will be disturbed.
Maintenance method of cm2600d spectrophotometer:
oldmeters.com can provide users with 3nh, Tilo, Ashley, Japan and other brand instrument maintenance services, which can solve the common problems such as unable to start, inaccurate measurement, battery range, key failure and so on. If your cm2600d spectrophotometer has the above common problems, you can contact our online customer service in time.