Home>Industry information

CR-10 Maintenance of color difference bulb

Release:999 instrument maintenance network time:2019-10-29

CR-10 is a small colorimeter produced by Japan. CR-10 is measured 2000 times by 5 batteries (× 4 cells, alkaline manganese battery: measured in 10 second interval; nickel cadmium battery: measured in 10 second interval, measured 600 times. )Power supply, measuring caliber 8mm, suitable for fast color control. CR-10 is a colorimeter designed by integrating measurement, data processing and display of three parts with 8 ° directional lighting diffusion receiving mode (including specular reflection light). CR-10 colorimeter has high cost performance and is widely used in rubber, plastic, interior decoration, clothing and other related industries.
 
 
 
 
 
Common faults of Japan CR-10 color difference instrument
 
 
 
Because the lamp life of this colorimeter is generally less than 200000 times, users who use CR-10 often find that the bulb is faulty. Replacing the bulb is the most common maintenance item.
 
 
 
 
 
Judgment of CR-10 bulb failure:
 
 
 
How to judge the problem of the CR-10 color difference lamp you purchased? When the power supply is turned on to measure the target, "illumination error" will be displayed on the instrument screen. If the power supply is pushed from the off position to the on position at this time and the error is still reported, and the error prompt is still displayed when the battery is replaced, it is likely that the CR-10 bulb has failed. The cause of the failure may be the filament is burnt out, and part of the measurement circuit may have failed.
 
 
 
 
 
What should I do if the CR-10 bulb fails?
 
 
 
The CR-10 bulb failure is a common problem, which can be solved by replacing the bulb. Therefore, it is recommended to contact us for maintenance and replacement after the failure, and we will replace the same type of CR-10 bulb for your colorimeter. After the bulb is replaced, we will calibrate the instrument to ensure that you can use the repaired instrument normally.
 
 
 
 
 
CR-10 colorimeter bulb failure
 