Home>Industry information

Analysis and maintenance of cm2500c common faults of colorim

Release:999 instrument maintenance network time:2019-10-29

Colorimeter cm2500c is a Portable Colorimeter introduced by Japan. Users may have common problems such as light source damage, system error reporting, unable to correct, etc. in the process of use. The following is a brief introduction to common fault analysis and maintenance methods of colorimeter cm2500c. Interested friends may wish to have a look!

cm2500c

Analysis of common faults of colorimeter cm2500c:
 
 
 
1. Color difference meter cm2500c shows "er05" flash failure. Possible causes:
 
 
 
Xenon lamp flickers abnormally, xenon lamp is broken, charging circuit is faulty, sensor is faulty.
 
 
 
2. The color difference meter cm2500c displays "general system error × A1". terms of settlement:
 
 
 
Clean the instrument and calibrate the whiteboard. Click here for cleaning instructions. Try to measure, if the error still occurs, continue to the next step.
 
 
 
Disconnect the power supply of the instrument, remove the battery, and then install it again in a few minutes.
 
 
 
Connect the power adapter and charge the battery for more than 1 hour.
 
 
 
Try to measure after charging. If the error persists, the battery may need to be replaced. If the battery is good or new and charged, proceed to the next step.
 
 
 
Restore factory settings. This will result in the loss of all data (standards and samples).
 
 
 
3. The color difference meter cm2500c shows "cannot be corrected". Possible causes:
 
 
 
The white correction may not have been carried out according to the correct procedure.
 
 
 
Please use the cleaned white correction plate to carry out the white correction again.
 
 
 
(for example) if you correct with something black, ER11 will appear.
 
 
 

If the corrective operation is normal, it may be a product failure colorimeter cm2500c RepairWay:

For such precision optical instruments as colorimeter, it may cause irreversible damage to the instrument if the instrument is disassembled without permission. In order to ensure the safety performance of the instrument, users need to select the official maintenance center of colorimeter brand or the authorized maintenance center for maintenance. oldmeters.com is a service platform jointly built by Shenzhen sanenchi Technology Co., Ltd. and Shenzhen sanenshi Technology Co., Ltd. for the maintenance of colorimeter, glossmeter and spectrophotometer. At present, it supports the maintenance of 3nh, Tilo, Japan, Ashley, BYK and other brand instruments to help users solve:
 
 
 
Screen does not display, flower screen
 
 
 
Battery life is short, battery is not durable
 
 
 
Measurement fault error
 
 
 
Main board water / no power on
 
 
 
Key is not working properly.
 
 
 
Correction failed
 
 
 
Unable to power on, screen does not light up
 
 
 
There are many problems, such as the instability of measurement data and the uncertainty of data.

Users only need to follow the chromatic aberration meter Maintenance process If you place an order, you can enjoy the convenient maintenance service of color difference meter.